Résumé en français · Traduction disponible
GB27955-2020
Hygienic requirements for low-temperature hydrogen peroxide gas plasma sterilizer
过氧化氢气体等离子体低温灭菌器卫生要求
GB 27955-2020 est une norme nationale chinoise obligatoire en vigueur : Hygienic requirements for low-temperature hydrogen peroxide gas plasma sterilizer. Secteur : Alimentation. Thème principal : Sécurité des produits. Entrée en vigueur : 1 nov. 2020.
L’applicabilité dépend du périmètre du produit, de la voie de certification, des avis des autorités et des règles de transition. Demandez une traduction ou vérifiez la source avant d’utiliser cette fiche pour une décision de conformité.
Aperçu de conformité
Cet aperçu n’est actuellement disponible qu’en anglais.
Champ d’application
This standard specifies the technical requirements and hygienic rules for low-temperature hydrogen peroxide gas plasma sterilizers. It applies to the sterilization of medical devices, instruments, and items that are not moisture-resistant and not high-temperature resistant. The standard explicitly prohibits the use of these sterilizers for liquids, powders, implants, and items with blind lumens. It references GB/T 1616 for industrial hydrogen peroxide quality.
Exigences techniques
The standard outlines the mandatory requirements for the sterilization process, which must include conditioning, sterilization, and ventilation stages. The sterilant used must be hydrogen peroxide with a concentration maintained between 53% and 60%. Manufacturers must ensure that the sterilizer displays and records critical parameters such as chamber temperature, pressure, and plasma input power. For sterilization efficacy, the standard requires a half-cycle full load test with no bacterial growth. Safety requirements cover environmental exposure limits, where the time-weighted average concentration of hydrogen peroxide must not exceed 1.5mg/m³. Additionally, the standard mandates biological compatibility and material compatibility to ensure no corrosion or adverse effects on sterilized items, referencing the GB/T 16886 series for biological evaluation.
Méthodes d’essai
Testing involves verifying the sterilization program parameters, including chamber pressure (≤80Pa), wall temperature (≤60°C), and plasma input power. Sterilization efficacy is evaluated using biological indicators, specifically Geobacillus stearothermophilus spores, placed in stainless steel and polytetrafluoroethylene (PTFE) lumens. Material compatibility is assessed by subjecting metal and non-metal samples to at least 100 sterilization cycles and evaluating corrosion rates. Environmental exposure testing follows GBZ/T 300.48.
Questions fréquentes
What products are regulated under GB 27955-2020?
It regulates low-temperature hydrogen peroxide gas plasma sterilizers used for heat- and moisture-sensitive medical devices.
What items cannot be sterilized under this standard?
The standard prohibits sterilizing liquids, powders, implants, items with blind lumens, and cellulose-based materials.
What is the required concentration of the hydrogen peroxide sterilant?
The hydrogen peroxide sterilant must have a concentration between 53% and 60%, complying with GB/T 1616.
The summary above provides a high-level overview of GB 27955-2020. To ensure complete compliance for your products entering the Chinese market, a precise and professional translation of the full technical specifications is essential. Purchase our high-quality English translation of GB 27955-2020 today to access all detailed parameters, tables, and testing procedures.
Indications d’applicabilité
- Type
- Produit
- CCS
- C50
- ICS
- 11.080
- Autorité compétente
- National Disease Control and Prevention Administration
- Comité technique
- National Disease Control and Prevention Administration
Relations
Remplace:GB 27955-2011